STABLO-AP N2 Ionizer - Static Electricity Remover in Nitrogen Gas
Brand: Shimadzu
World’s First Static Charge Removal in Nitrogen Gas Environments
STABLO-AP N2 Ionizer - Static Electricity Remover in Nitrogen Gas – Overview
World’s First Static Charge Removal in Nitrogen Gas Environments
STABLO-AP N2 Ionizer - Static Electricity Remover in Nitrogen Gas – Specs
STABLO-AP Key Specifications
| Model Name | STABLO-AP | STABLO-AP Ar | STABLO-AP N₂ | STABLO-AP DRY AIR |
|---|---|---|---|---|
| Operating Environment*1 | Typical atmospheric air environment Temp: 0 to 40 °C Humidity: 25 to max. 85 %RH (with no condensation) | Argon gas environment*2,*3 Temp: 0 to 40 °C Dew Point: −76 °Cdp (Moisture Concentration 1 ppm) Oxygen Concentration: 1 ppm | Nitrogen gas environment*2,*3,*4,*5 Temp: 0 to 40 °C Dew Point: −76 °Cdp (Moisture Concentration 1 ppm) Oxygen Concentration: 1 ppm | Dry air environment*2,*3,*4 Temp: 0 to 40 °C Dew Point: −30 °Cdp (Moisture Concentration: 375 ppm) |
| Static Elimination Method | AC corona discharge | AC corona discharge | AC corona discharge | AC corona discharge |
| Ion Balance*6,*7 | ±10 V | ±10 V | ±10 V | ±10 V (Dew Point: −70 °Cdp) ±30 V (Whole operating environment) |
| Effective Static Removal Range*7 | Approx. 300 mm from the outlet | Approx. 300 mm from the outlet | Approx. 300 mm from the outlet | Approx. 300 mm from the outlet |
| Static Elimination Time*6,*7,*8 | Approx. 1 sec. | Approx. 2.5 sec. | Approx. 5 sec. | Approx. 3.5 sec. |
| Ozone Concentration | 0.06 ppm max. (100 mm from discharge port) | Not applicable*9 | Not applicable*9 | 0.06 ppm max.*10,*11 (100 mm from discharge port) |
| Discharge Electrode (Material) | Tungsten (durability: 30,000 hours) | Tungsten (durability: 30,000 hours) | Tungsten (durability: 30,000 hours) | Tungsten (durability: 30,000 hours) |
| Size (unit) | Approx. 120 × 87 × 49 mm (Excluding protruding parts) | Approx. 120 × 87 × 49 mm (Excluding protruding parts) | Approx. 120 × 87 × 49 mm (Excluding protruding parts) | Approx. 120 × 87 × 49 mm (Excluding protruding parts) |
| Weight | Approx. 710 g (Main unit: 395 g, Stand: 315 g) | Approx. 710 g (Main unit: 395 g, Stand: 315 g) | Approx. 710 g (Main unit: 395 g, Stand: 315 g) | Approx. 710 g (Main unit: 395 g, Stand: 315 g) |
| Rated Electric Power Supply | DC 24 V, 0.1 A | DC 24 V, 0.1 A | DC 24 V, 0.1 A | DC 24 V, 0.1 A |
| Input Power Supply | AC 100 V–240 V, 0.58 A, 50/60 Hz*12 | AC 100 V–240 V, 0.58 A, 50/60 Hz*12 | AC 100 V–240 V, 0.58 A, 50/60 Hz*12 | AC 100 V–240 V, 0.58 A, 50/60 Hz*12 |
*1 It features an auto-off function to reduce wear on the discharge needle
(automatically stops 15 minutes after ion irradiation).
*2 To achieve the specified performance, allow the ionizer to acclimate in the
operating environment for at least 24 hours to ensure moisture dissipates from
the ionizer body.
*3 Pressure reduction in the glove box pass box is permissible down to −0.1 MPa.
*4 Do not place any conductive materials around the ionizer or in the space
between the ionizer and the object for static elimination. The ions may be
attracted, causing a deterioration in static elimination performance.
*5 When installing a nitrogen gas environment model in an AP-AD/AP series
balance, we recommend extending the ion irradiation time setting from
10 seconds to 30 seconds.
*6 This shipping inspection criterion is measured with a 20 pF 150 × 150 mm
CPM (charged plate monitor) plate placed at a 100 mm distance from the center
of the outlet.
*7 When using the ionizer in an environment with airflow, such as a purge-type
glove box, make sure to direct the ionizer outlet along the direction of the airflow.
*8 Time for 1 kV of static charge to decrease to 100 V or less
(at 100 mm distance).
*9 STABLO-AP Ar and N₂ do not generate ozone, because oxygen does not exist
in an Argon gas and Nitrogen gas environment.
*10 Please ensure adequate ventilation, as ozone concentrations may increase
near the irradiation port (within 50 mm) and in enclosed spaces, which may
cause irritation to the nose and throat.
*11 When using it installed on an AP-AD/AP series balance, the ozone
concentration will be higher inside the draft shield due to the enclosed space.
Please leave the ion irradiation time at the default setting of 10 seconds,
and do not perform continuous ion irradiation for more than 10 seconds.